Turbospec Analizer
Frequently Asked Questions

icon

Turbospec quantitative elemental analyzer incorporates industry-leading operational and performance features. The analyzer embodies the mission of the company by offering operational simplicity with unprecedented accuracy and time to results. We share a list of Frequently Asked Question, which was drafted after closely working with our customers in all continents.

The Turbospec e400

There is a long list of NMR-active elements that may be measured by the Turbospec analyzers. Turbospec has created a list of high-precision elements of interest, principally driven by demand in mining and food applications. 

Here are the most commonly offered elements:

  • Sodium
  • Lithium
  • Boron
  • Phosphorous
  • Proton
  • Fluorine
  • Chlorine. 

Please inquire about the different elements required by specific applications.

Liquid samples are measured as-is.

Filtration is necessary only to remove solid particles.

If you start with a solid sample, read the explanation below.

The Turbospec analyzer measures the number of atoms per unit of volume in a solution by measuring the amplitude of Nuclear Magnetic Resonance signals.

As the instrument uses radio waves to inspect the sample, the measurement is not affected by the transparency or cleanliness of the solution – in contrast to optical techniques such as IR.

The Turbospec technology has several advantages over titration methods:

  • No consumables.
  • The liquid sample is measured as is.
  • There is very minimal training required.
  • High accuracy due to no sample preparation errors.

The Turbospec analyzers costs significantly less than ICP and most high-end AA units. Maintenance and calibration are remarkably simple. 

The list of elements to be analyzed by Turbospec is shorter than the ICP list. Therefore, Turbospec shows value when immediate accurate results are necessary for the list of elements offered for the analysis, mainly Sodium, Lithium, Chlorine, and Boron. 

ICP and AA require solutions clear of unwanted elements and significant sample dilution. These are non-issues for the Turbospec technology. This simplifies the laboratory operation and reduces sources of systematic errors. 

No. Because the measurement is based on the response of the atoms excited during the NMR scan, other elements in the sample do not alter the readings. The instruments selcorrects for matrix effetcs, present in high TDS solutions. 

Yes, solid samples can be analyzed, but only after the element of interest is in a solution. For example, to measure the sodium concentration in potato chips, the sample is crushed, water is added, and the measurement is done in the semi-liquid sample. 

For rock samples, methods such as microwave digestion need to be used to have the mineral samples turned into a solution.

Turbospec has experience in a number of applications and will provide guidance when the samples are not in liquid form.

Only large solid material from the sample – if present – needs to be removed.

For example, to measure the concentration of sodium in soup, liquid may be pipetted directly from the liquid stock and placed into the measurement tube. Or for total sodium content, the soup should be diluted and blended prior to placing the liquid in the NMR tube. 

The precision depends on the specific element. For example, for Lithium and Boron, the precision is 5 to 7 ppm (parts per million in relative weight).

The Limit of Detection (LOD) is approximately 3 times the standard deviation of the measurement. The Limit of Quantification (LOQ) is 10 times the standard deviation of the measurement.

The LOD for elements such as Lithium and Boron is approximately 15 ppm. Custom programming is offered for applications requiring lower LOD.

Only 500 uL (0.5 ml) of solutionos needed. The 5 mm diameter glass tube needs to be filled to 40 mm. 

Instruments are factory-calibrated. Customers may choose to overwrite the factory calibration using their own standards for a single-point calibration. 

Yes. Turbospec includes up to six elements per instrument.

The most common required elements in the Lithium industry are Sodium, Lithium, Chlorine, and Boron. In the food industry, Sodium, Chlorine, and Phosphorous.

The liquid samples are measured as is. Dilution and blending are necessary for samples such as soups and potato chips. 

The standard samples may be a solution with accurate and known concentration or may be purchased from a chemical company or prepared by the client.

The main advantages of the Turbospec technology over conventional analytical instruments is that the sample is measured as is and the calibration and measurement are remarkably simple. There is no need for consumables as glass tubes may be reused. 

This means that there are no systematic errors associated with dilution and filtration steps. 

NMR offers a more direct measurement of the concentration of elements in solution.

A direct comparison with other techniques needs to take into consideration the systematic errors associated with sample preparation, in particular dilution.

The Turbospec reading is less prone to operator errors due to the simplicity of the protocols.

The Turbospec analyzers can be purchased or rented. The minimal rental term is three months.

No consumables are needed to operate the Turbospec methods. Glass tubes may be reused. Some clients choose to buy a stock of glass tubes so they don’t need to clean used tubes frequently. 

No. The instrument is plug and play. The simple installation safety considerations and protocols are presented in the User Guide.

5 mm diameter NMR glass tubes. 4” length is recommended. 7” long tubes may also be used. 

Size: 48 cm width, 25 cm height, 25 cm depth

Weight: 8.3 Kg

Request a quotation from sales@turbospecllc.com. Indicate the application, elements of interest, and typical concentrations to be measured.

The Turbospec units are shipped inside a custom-hardened case with laser-cut foam. The case has wheels.

For domestic deliveries, the instrument is shipped in the case.

For international deliveries, the case is placed inside a custom wooden crate – certified for freight.

For precision readings, it is recommended to use the Turbospec units indoors.

The instrument may be used outdoors or inside a vehicle (request custom cable).

Avoid extended sun exposure and ingress of liquid or dust from the rear vents.

Power: input 10 to 28 V DC at 5 A. 110-240 V AC transformer and power cord included.

Simply send a message with your support needs to support@turbospecllc.com.

Selecting the PREVIOUS RESULTS tab at the bottom of the screen displays the stored list of previous results.

The display shows the date and time of the measurement, the type of entry (measurement result or calibration), the value, and any user-entered notes. Up/Down finger swipes will scroll the display. 

Custom notes may be added to the readings: After a measurement, the user can add a notation in the results file by selecting the ADD NOTES tab at the bottom of the screen.

The note may be up to 30 characters long. Notes can only be added for a measurement immediately after that measurement is made.

Yes. Once a USB drive is plugged in any of the three USB outlets, a “SAVE TO FLASH” message appears in the Previous Results window. Press the bottom. Data is saved in a format that can be opened as a text file or in Excel. 

Inserting a Sample: 

Slide the 5 mm tube down until it reaches the stop. 

Calibrating to overwrite the factory calibration): 

  • Select a calibration reference and insert it into the analyzer.
  • Press the CALIBRATION button on the right of the display.
  • Use the numerical keyboard that appears on the screen to enter the concentration value you associate with this calibration standard. Molarity is the recommended unit to be used for calibration. You may use other units of concentration, e.g. mg/ml or ppm. 
  • Press OK to begin the calibration. When the calibration process complete, the display will indicate that it is ready for measurements.

Measuring Concentration: 

Insert the sample tube into the sample holder and press the SCAN button. The element concentration value will be displayed as soon as the scan is complete. The units are the same ones used during calibration. Correct for density if you use ppm.

Common solute concentration units used for elemental analysis are Molarity and parts per million. Molarity is the preferred unit to be used in the Turbospec measurements, as the analyzer measures the number of atoms per unit volume.

Molarity (M) = number of moles of solute/liter of solution

The Molarity is proportional to the number of atoms per unit of volume.

Other commonly used units are parts per million:

Parts per million (ppm) = (mass of solute/mass of solution) x 106

As the concentration in ppm is the number of grams of solute per gram of solution times 106, a density correction is necessary when converting units from ppm to M. This correction is not needed for dilute solutions but is relevant when the density of the solution differs significantly from 1 gr/mL.

To convert from concentration in ppm (C[ppm]) to Molarity (C[M]) use the following formula:

C[ppm] = C[M] x (atomic weight/density of solvent) x 1000

For example, in an 8 M solution of LiCl with a density of 1.15 gr/mL, the Lithium concentration in ppm is:

C[ppm] = 8 M x (6.941 gr/M /1.15 gr/mL) x 1000 = 48,285 (mg of Lithium per kg of solvent).

  • Size: 48 cm width, 25 cm height, 25 cm depth
  • Weight: 8.3 Kg
  • Power: input 10 to 28 V DC at 5 A. 110-240 V AC transformer and power cord included
  • Temperature of Operation: 10 C – 50 C
  • Exchangeable fuse
  • RFB Radio-Frequency Controller: single channel broadband ADC reprogrammable system
  • Connectivity: USB and Ethernet
  • Magnet: ultra-compact high homogeneity permanent magnet with passive shielding for stray fields
  • Touch screen configuration
  • Touch Screen: 8″ TFT HDMI high-brightness LCD
  • Cooling: fan in back panel
  • User Interface: Graphical User Interface with intuitive operation
  • Sample Vial: glass test tube, 5 mm outside diameter, 100 mm length or longer

Secure your order by placing the Purchase Order and paying the deposit. Confirm the lead time from Turbospec’s Sales Team. Coordinate the shipment with your preferred freight company.

The customer is responsible for the payment of taxes and shipping arrangements, including customs duties. If you need support with the shipment, Turbospec will provide expert help and all the necessary documentation.